pHionics 4-20 mA pH Sensor Calibration Instructions
-
- Rinse electrode with distilled water.
- Submerse the sensor in 7 pH buffer, letting the sensor and calibration solution equilibrate to the same temperature. Wait for mA output to stabilize.
-
- Reading should be 12.00 ± 0.60 mA for 7 pH buffer. If it is not, the electrode may require cleaning. Follow the instructions in section 4.2 for proper procedures as they are the same as dissolved oxygen electrodes.
- Always ensure solution is in contact with the metal housing for proper readings.
-
- Record mA output for 7 pH buffer.
- Rinse electrode with distilled water.
- Submerse sensor in secondary calibration buffer and wait for reading to stabilize.
-
- Reading should be 8.56 ± 0.60 mA for 4 pH and 15.4 ± 0.60 mA for 10 pH buffer.
- 10 pH buffer is known to be unstable and should always be prepared fresh.
-
- Record mA output for secondary buffer and input into software if using automatic scaling, or use the calculator below to generate a linear equation using the two recorded data points.
- Rinse electrode with distilled water.
- The sensor is now ready for use.
- Always keep electrode tip in saturated potassium chloride solution when not in use.
Calibration is now complete. Follow this procedure anytime calibration of the sensor is required.
4–20 mA to pH Scaling Calculator
Enter two points from your sensor’s range to get the linear equation for your PLC.
Calibration tip: Readings between your two calibration points (interpolation) are more accurate than readings beyond them (extrapolation). Use the two buffers that most closely bracket your expected sample pH. For example, use pH 4 and 7 buffers if the sample liquid is typically between 5 and 6 pH. Use pH 7 and 10 buffers if the sample is typically between 7 and 9 pH.
y = pH, x = mA | slope (m) = , intercept (b) =