Click the following links to jump to your specifications of interest:

STs4xxx -- for submersion applications, 4-20 ma output
STs4xxxT -- for submersion applications with additional temperature output (4-20 ma)
STs4xxxB -- for submersion applications, 0.40 to 2.50 vdc output, battery powered
How to specify range
How to specify custom range
download STs4xxx conductivity brochure (130kb)? STs4xxx conductivity water quality transmitter PDF

 
series STs4xxx, 2-wire, 4-20 ma conductivity sensor/transmitters

output

4 to 20 ma

power supply voltage

7 to 40 vdc

loop impedance (max)

250 ohms at 12 vdc,  1650 ohms at 40 vdc

cable from transmitter to power supply

2 conductor, twisted pair, 3 mile maximum

isolation

600 VDC, >70 db at 50/60 Hz

See additional specifications below that apply to all versions -- click here

to top

 

 

series STs4xxxB, voltage output conductivity sensor/transmitters replaceable battery, or for external supply provided by user

output

0.40 to 2.50 volts (or specify desired output)

power supply voltage

self contained, replaceable lithium battery, or external 4 to 16 volts DC supply 

output impedance (max)

1 Megohm (prevents premature failure of battery if leads are shorted)

cable from transmitter to power supply

4 conductor, twisted pair, 1000 feet max

isolation

none

See additional specifications below that apply to all versions -- click here

to top

 

 

series STs4xxxT, 2-Wire, 4-20 ma conductivity sensor/transmitters with
optional,
additional 2-wire, 4-20 ma temperature output

output

4 to 20 ma

range

0-50 degrees Celsius

power supply voltage

7 to 40 vdc

loop impedance (max)

250 ohms at 12 vdc, 1650 ohms at 40 vdc

cable from transmitter to power supply

4 conductor, twisted pair, 3 mile maximum

isolation

600 vdc, >70 db at 50/60 Hz

See additional specifications below that apply to all versions -- click here

to top

 

 

The following data pertains to all configurations:

linearity

± 1.0% of full scale

accuracy

± 1.0% of full scale

sensitivity

± 0.05% of full scale

stability

± 0.1% of full scale

repeatability

± 0.1% of full scale

response time (including electrodes)

90% < 5 seconds

temperature compensation

2% per degree Celsius 

input range

0-100, 200, 500, 1000, 2000, 5000, 10000 uSiemens see below

conductivity sensing range

0-100, 200, 500, 1000, 2000, 5000, 10000 uSiemens see below

pressure

0-100 psi

humidity

0-100%

wetted materials

316 SS, Kynar, Teflon, Viton
optional titanium housing

length

343 mm  (13.5 in.)

diameter

19 mm  (0.750 in.)  maximum

standard cable length, model STs 10 meters (approx 33 feet)

weight (excluding cable)

< 0.22 kg   (0.5 lb.)

to top

 

 

How to specifiy range:  Replace the xxx model number with the corresponding range desired:
range model number
0-100 uS   STs4102 or STs4102T          (1.0 x 102)
0-200 uS   STs4202 or STs4202T          (2.0 x 102)
0-500 uS   STs4502 or STs4502T          (5.0 x 102)
0-1000 uS   STs4103 or STs4103T          (1.0 x 103)
0-2000 uS   STs4203 or STs4203T          (2.0 x 103)
0-5000 uS   STs4503 or STs4503T          (5.0 x 103)
0-10000 uS    STs4104 or STs4104T          (1.0 x 104)
0-specify uS    STs4xxx or STs4xxxT Replace xxx with appropriate numbers -- see examples below

to top

 

 

To determine your custom range, observe the following examples or contact pHionics at the numbers listed below:
desired range   decimal notation   appropriate model number
0-250   2.5 x 102 STs4252 or STs4252T
0-2500   2.5 x 103 STs4253 or STs4253T
0-300 3.0 x 102 STs4302 or STs4302T
0-3000 3.0 x 103 STs4303 or STs4303T

to top

 

home  contact us  how do I order?  discount policy  oem's  links   about pHionics  