conductivity
specifications
Click the
following links to jump to your specifications of interest:
STs4xxx
-- for submersion applications, 4-20 ma output
STs4xxxT
--
for submersion applications with additional temperature output (4-20 ma)
STs4xxxB
-- for submersion applications, 0.40 to 2.50 vdc output, battery powered
How
to specify range
How
to specify custom range
download
STs4xxx conductivity
brochure (130kb)?

|
|
|
series
STs4xxx,
2-wire, 4-20 ma conductivity sensor/transmitters
|
|
output
|
4 to 20 ma
|
|
power supply voltage
|
7 to 40 vdc
|
|
loop impedance (max)
|
250 ohms at 12 vdc,
1650 ohms at 40 vdc
|
|
cable from transmitter to
power supply
|
2 conductor, twisted
pair, 3 mile maximum
|
|
isolation
|
600 VDC, >70 db at
50/60 Hz
|
|
See
additional specifications below that apply to all versions -- click
here
|
|
to
top
|
|
|
|
series
STs4xxxB,
voltage output conductivity sensor/transmitters replaceable battery, or for external supply provided by user
|
|
output
|
0.40 to 2.50 volts (or
specify desired output)
|
|
power supply voltage
|
self contained,
replaceable lithium battery, or external 4 to 16 volts DC supply
|
|
output impedance (max)
|
1 Megohm (prevents
premature failure of battery if leads are shorted)
|
|
cable from transmitter to
power supply
|
4 conductor, twisted
pair, 1000 feet max
|
|
isolation
|
none
|
|
See
additional specifications below that apply to all versions -- click
here
|
|
to
top
|
|
|
|
series
STs4xxxT,
2-Wire, 4-20 ma conductivity sensor/transmitters with
optional, additional 2-wire, 4-20 ma temperature output
|
|
output
|
4 to 20 ma
|
|
range
|
0-50 degrees Celsius
|
|
power supply voltage
|
7 to 40 vdc
|
|
loop impedance (max)
|
250 ohms at 12 vdc, 1650
ohms at 40 vdc
|
|
cable from transmitter to
power supply
|
4 conductor, twisted
pair, 3 mile maximum
|
|
isolation
|
600 vdc, >70 db at
50/60 Hz
|
|
See
additional specifications below that apply to all versions -- click
here
|
|
to
top
|
|
|
|
The following
data pertains to all configurations:
|
|
linearity
|
± 1.0% of full scale
|
|
accuracy
|
± 1.0% of full scale
|
|
sensitivity
|
± 0.05% of full scale
|
|
stability
|
± 0.1% of full scale
|
|
repeatability
|
± 0.1% of full scale
|
|
response time (including
electrodes)
|
90% < 5 seconds
|
|
temperature compensation
|
2% per degree Celsius
|
|
input range
|
0-100, 200, 500, 1000,
2000, 5000, 10000
uSiemens see below
|
|
conductivity sensing
range
|
0-100, 200, 500, 1000,
2000, 5000, 10000
uSiemens see below
|
|
pressure
|
0-100 psi
|
|
humidity
|
0-100%
|
|
wetted materials
|
316 SS, Kynar, Teflon,
Viton
optional titanium housing
|
|
length
|
343 mm (13.5 in.)
|
|
diameter
|
19 mm (0.750 in.)
maximum
|
| standard
cable length, model STs |
10
meters (approx 33 feet) |
|
weight
(excluding cable)
|
< 0.22 kg (0.5
lb.)
|
|
to
top
|
|
|
| How
to specifiy range: Replace
the xxx
model number with the
corresponding range desired:
|
| range |
model
number |
| 0-100 uS |
STs4102
or STs4102T
(1.0 x 102) |
| 0-200 uS |
STs4202
or STs4202T
(2.0
x 102) |
| 0-500 uS |
STs4502
or STs4502T
(5.0 x
102) |
| 0-1000 uS |
STs4103
or STs4103T
(1.0
x 103) |
| 0-2000 uS |
STs4203
or STs4203T
(2.0
x 103) |
| 0-5000 uS |
STs4503
or STs4503T
(5.0
x 103) |
| 0-10000 uS
|
STs4104
or STs4104T
(1.0
x 104) |
| 0-specify uS
|
STs4xxx
or STs4xxxT
Replace xxx with
appropriate numbers -- see examples below |
|
to
top |
|
|
| To determine your custom
range, observe the following examples or contact pHionics at the numbers listed below: |
| desired
range |
decimal
notation |
appropriate
model number |
| 0-250
|
2.5
x 102 |
STs4252
or STs4252T |
| 0-2500
|
2.5
x 103 |
STs4253
or STs4253T |
| 0-300 |
3.0
x 102 |
STs4302
or STs4302T |
| 0-3000 |
3.0
x 103 |
STs4303
or STs4303T |
|
to
top |
|
|
home
contact
us
how do I order?
discount policy
oem's
links
about pHionics
wire transfers
conductivity
pH
orp (redox)
dissolved oxygen
temperature
voice:
1-800-964-0063 (USA),
1-775-246-2711 fax:
1-775-246-2722
e-mail:
contact phionics
web:
http://www.phionics.com
©